Reference intervals for the white blood cell and differential white blood cell count in the adult population at the use of the automatic hematology analyzer Sysmex XT 2000i
https://doi.org/10.18821/0234-5730-2016-61-3-153-156
Abstract
Modern international standards recommend to each laboratory to develop or to confirm the reference intervals for laboratory parameters. The purpose of the study was to establish reference intervals for the quantity of white blood cell and differential white blood cell count in the adult population of the Astrakhan region at the use of the automatic hematology analyzer Sysmex XT 2000i. In the group of 375 healthy people there were determined the white blood cell and differential white blood cell count by the flow cytometry and the cytofluorometry. The reference intervals for the white blood cell and differential white blood cell count didn’t depend on a gender.
About the Authors
O. B. PetrovaRussian Federation
Petrova Olga V., MD., PhD., head of the laboratory
Astrakhan, 414011
G. R. Shabanova
Russian Federation
Astrakhan, 414011
T. G. Egorova
Russian Federation
Astrakhan, 414011
References
1. Alan H.B. Tietz clinical guide to laboratory tests. Мoscow: Labora; 2013. (in Russian)
2. Kazakova M.S., Lugovskaya S.A., Dolgov V.V. Reference interval of indicators of the total blood analysis of the adult working population. Clinical laboratory diagnostics. Russian journal (Klinicheskaya laboratornaya diagnostika). 2012; 6: 43–9. (in Russian)
3. Kishkun A.A. Guide to laboratory methods of diagnostics. Moscow: GEOTAR-media; 2007. (in Russian)
4. Petrova O.V., Shashin S.A., Tarasov D.G. Value of unripe granulocytes in diagnostics of infectious and inflammatory processes at cardiac patients. Clinical laboratory diagnostics. Russian journal (Klinicheskaya laboratornaya diagnostika). 2014; 5: 25–40. (in Russian)
5. Shaw J.L., Cohen A., Konforte D., Binesh-Marvasti T., Colantonio D.A., Adeli K. Validity of establishing pediatric reference intervals based on hospital patient data: a comparison of the modified Hoffmann approach to CALIPER reference intervals obtained in healthy children. Clin. Biochem.2014; 47(3): 166–72. doi: 10.1016/j.clinbiochem.2013.11.008.
6. Ridefelt P., Hellberg D., Aldrimer M., Gustafsson J. Estimating reliable pediatrics reference intervals in clinical chemistry and haematology. Acta Paediatr.2014; 103(1): 10–5. doi: 10.1111/apa.12438.
7. Sinclair L., Hall S., Badrick T. A survey of Australian haematology reference intervals. Pathology. 2014; 46(6): 538–43. doi: 10.1097/ PAT.0000000000000148.
8. Bosco Gahutu J. Clinical chemistry reference intervals in a Rwandan population. Br. J. Medicine Medical Res. 2013; 3(3): 532–42. doi: 10.9734/bjmmr/2013/2741.
9. Henry E., Christensen R.D. Reference intervals in neonatal hematology. Clin. Perinatol. 2015; 42(3): 483–97. doi: 10.1016/j.clp.2015.04.005.
10. Zierk J., Arzideh A., Haeckel R., Rascher W., Rauh M., Metzler M. Indirect determination of pediatric blood count reference intervals. Clin. Chem. Lab. Med. 2013; 51(4): 863–72. doi: 10.1515/cclm-2012-0684.
11. Defining, establishing, and verifying reference intervals in the clinical laboratory: approved guideline. CLSI C28-A3. Wayne: Clinical Laboratory Standards Institute; 2008. Available at: http://www.clsi.org
12. Bertholf R.L. Statistical methods for establishing and validating reference intervals. Lab. Medicine. 2006; 37(5): 306–10. doi:10.1309/cbmn-prfnlu1x-a4xv.
13. Bolann B.J. Easy verification of clinical chemistry reference intervals. Clin. Chem. Lab. Med. 2013; 51(11): e279–81. doi:10.1515/cclm-2013-0356.
14. Daly C.H., Liu X., Grey V.L., Hamid J.S. A systematic review of statistical methods used in constructing pediatric reference intervals. Clin. Biochem. 2013; 46(13–14): 1220–27. doi:10.1016/j.clinbiochem.2013.05.058.
15. Blankenstein M.A. Reference intervals – eves met a normal person? Ann. Clin. Biochem. 2015; 52(Pt 1): 5–6. doi: 10.1177/0004563214561503. Available at: http://www.acb.sagepub.com
16. Katayev A., Balciza C., Seccombe D.W. Establishing reference intervals for clinical laboratory test results: is there a better way? Am. J. Clin. Pathol. 2010; 133(2): 180–6. doi: 10.1309/ajcpn5bmtsf1cdyp.
17. Horowitz G.L. Estimating reference intervals. Am. J. Clin. Pathol. 2010; 133(2): 175–7. doi: 10.1309/AJCPQ4N7BRZQVHAL.
18. Petrova O.V., Tarasov D.G., Zakharova L.R. Experience of introduction of LIS PSM in cliniko-diagnostic laboratory. Poliklinika. 2011; 4(2): 36–7. (in Russian)
19. Melzer S., Zachariae S., Bocsi J., Engel C., Loffler M., Tarnok A. Reference intervals for leukocyte subsets in adults: Results from a populationbased study using 10-color flow cytometry. Cytometry B Clin. Cytom. 2015; 88(4): 270–81. doi: 10.1002/cyto.b.21234.
20. Lim E.M., Cembrowski G., Cembrowski M., Clarke G. Race-specific WBC and neutrophil count reference intervals. Int. J. Lab. Hematol. 2010; 32(6, Pt 2); 590–7. doi: 10.1111/j.1751-553X.2010.01223.x.
21. Kazakova M.S., Lugovskaya S.A. Reference interval of indicators of the total blood analysis at use of the automatic hematologic analyzer. Clinical laboratory diagnostics. Russian journal (Klinicheskaya laboratornaya diagnostika). 2014; 9: 10–1. (in Russian)
Review
For citations:
Petrova O.B., Shabanova G.R., Egorova T.G. Reference intervals for the white blood cell and differential white blood cell count in the adult population at the use of the automatic hematology analyzer Sysmex XT 2000i. Russian journal of hematology and transfusiology. 2016;61(3):153-156. (In Russ.) https://doi.org/10.18821/0234-5730-2016-61-3-153-156